Aptamer-Based Microcantilever Biosensors for Label-Free PSA Detection

Fei Gao,Zhao Qi Li,Jiao Yang Li,Zhe Yi Li,Jie Wei,Xumin Ding,Lei Wang,Lei Ying Zhai,Jing Chang Nan,Cong Wang
DOI: https://doi.org/10.1109/jsen.2024.3496092
IF: 4.3
2024-01-01
IEEE Sensors Journal
Abstract:Single-ended clamped microcantilever biosensors and qual-ended clamped with serpentine cantilevers microcantilever biosensors were developed and batch-fabricated using micro-electro-mechanical system technology for accurately measuring prostate-specific antigen (PSA) via a laser emitted by the confocal fiber displacement sensor. The aptamer was immobilized on the gold surface of the sensing cantilever so that interactions between the aptamer and PSA successfully generated a stress change of the microcantilever biosensors, resulting in a bending conformation. The reference cantilever without gold eliminates interference from the environment by detecting deflection induced by nonspecific interactions. The profile of microcantilevers could be monitored by a spectral confocal three-dimensional (3D) morphological scanning and imaging system. The two microcantilever deflections had a good linearity within 50 to 2000 ng/mL. The detection limits of the two microcantilever are 9.88 ng/mL and 7.08 ng/mL, respectively. These microcantilever biosensors provide a promising approach to detect PSA in early prostate cancer diagnosis.
What problem does this paper attempt to address?