Non-contact Monitoring of Human Cardiorespiratory Activity During Sleep Using FMCW Millimeter Wave Radar

En-Kang Wu,Qi-Gao Fan,Ming-Chao Li,Ji-Hao Zhang,Jie Jia,Tian Qiang,Cong Wang,Xiao-Feng Gu,Jun-Ge Liang
DOI: https://doi.org/10.1016/j.measurement.2024.116144
IF: 5.6
2025-01-01
Measurement
Abstract:Millimeter wave radar can be utilized for non-contact, long-term, and accurate health monitoring. This paper presents an application for vital signs detection including respiratory rate (RR) and heart rate (HR) during sleep, as well as facilitating rapid detection and extended recording of apnea syndrome. This paper employs frequency spectrum analysis for RR estimation and develops an algorithm of template matching estimation (TME) for real-time, accurate HR estimation by comparing radar signal fragments to a template with a frequency signature. The time required for HR estimation is reduced to 1 s. In addition, a neural network-based model was developed for detecting the apnea events with the accuracy of 93.60% and calculates the duration in seconds. The system underwent testing using irregular analog signals of vital signs. In the actual scene testing, the accuracy of the RR estimation is 90.58%, and 90.25% for HR estimation.
What problem does this paper attempt to address?