Clinical application of improved 3D printing navigation guiding template for atlantoaxial pedicle screw placement
王玉强,牛慧霞,赵耀,刘屹林,张敏,刘宏建,张书豪,赵亮,宋瑞鹏,王卫东,朱忠培,王利民
DOI: https://doi.org/10.3760/cma.j.cn421213-20190929-01405
2020-01-01
Abstract:Objective:To investigate the operative and clinical performance of an improved 3D printed navigation guiding template to develop patient specific solutions for posterior atlantoaxial pedicle screw fixation surgery.Methods:Thirty-eight patients with atlantoaxial dislocation admitted between January 2015 to December 2018 were divided into a template navigation group (
n=18) and a traditional group (
n=20). The template navigation group was assisted by the improved digital guide board, and the traditional group was guided by the C-arm fluoroscopy. The operative time, C-arm fluoroscopy times, blood loss and other intraoperative indicators were compared between two groups. Kawaguchi method was used to evaluate the satisfaction of screw placement. The accuracy and safety of the screw insertion were evaluated using preoperative and postoperative computed tomographic scan by calculation of screw deviation from the preplanned trajectory and evaluation of screw entry point location. Japanese Orthopedic Association scale (JOA) score and Neck Disability Index (NDI) were used to evaluate the improvement of postoperative symptoms in the two groups. Independent sample
t test was used for evaluation of intergroup patient age, follow-up time, efficacy. Chi-square test was used for comparison of counting data between the two groups, and paired
t test was used for comparison of measurement data within the group.
Results:All patients underwent surgery successfully. A total of 70 pedicle screws were inserted in the template navigation group, while 75 pedicle screws were inserted in the traditional group. The operative time (87.6±35.4) and frequency of fluoroscopy (3.9±1.9) in the template navigation group were significantly reduced as compared with those in the traditional group [Operative time (124.3±56.6) min (
χ2=3.734,
P<0.05), fluoroscopy frequency (12.4±4.7) (
χ2=4.327,
P<0.05)]. The amount of bleeding [(84.5±55.2) ml] in the template navigation group was slightly les than that in the traditional group [(96.1±69.4) ml], but the difference between the two groups was not statistically significant (
χ2=0.875,
P>0.05). JOA scores and NDI scores in both groups were improved similarly at each time point during the follow-up period. However, postoperative CT showed that the total accuracy rate of Kawaguchi 0-grade in template navigation group was 95.7% (67/70), significantly higher than that in the traditional group [77.3% (58/75),
χ2=2.628,
P<0.05]. In the template navigation group, there were no statistically significant differences in the actual measured inclination angle, head inclination angle, and the coordinate position of the nailing point before and after surgery, while there were statistically significant differences in the traditional group.
Conclusion:The improved digital navigation template is simple and safe to be applied in placement of atlantoaxial pedicle screws, and it can significantly shorten the operative time and reduce radiation exposure. Furthermore, the clinical effect of these patients is satisfactory.