Robotic Assistance for Upper Cervical Instrumentation: Report on Accuracy and Safety

Kaihang Luo,Xuenong Zou,Wei Chen,Shangbin Cui,Shaoyu Liu,Liuyun Chen,Lin Zhou
DOI: https://doi.org/10.1007/s00586-024-08510-4
IF: 2.721
2024-01-01
European Spine Journal
Abstract:This study aims to evaluate the accuracy and safety of C1 lateral mass and upper cervical pedicle screw placement assisted by the TiRobot II system. Ten patients who underwent cervical spine surgery assisted by the TiRobot II system were included. Screw accuracy was assessed using the Gertzbein-Robbins scale and by comparing the final screw positions with pre-planned trajectories. Deviations in screw tip, tail, and angle were recorded. Clinical data, including symptoms, surgical outcomes, and postoperative follow-up, were collected. Neurological improvement was evaluated using pre- and post-operative mJOA scores, with recovery rates calculated by Hirabayashi's method to assess outcomes 3 months after surgery. A total of 30 screws were placed in 10 patients. All screws (30/30) were within the clinically acceptable range, with 93.33
What problem does this paper attempt to address?