Roughhousing with Ions: Surface-Induced Dissociation and Electron Capture Dissociation as Diagnostics of Q-Cyclic IMS-TOF Instrument Tuning Gentleness

Andrew Arslanian,Vicki Wysocki
DOI: https://doi.org/10.26434/chemrxiv-2024-jfjt1
2024-11-05
Abstract:Native mass spectrometry can characterize a range of biomolecular features pertinent to structural biology, including intact mass, stoichiometry, ligand-bound states, and topology. However, when an instrument’s ionization source is tuned to max-imize signal intensity or adduct removal, it is possible that the biomolecular complex’s tertiary and quaternary structures can be rearranged in a way that no longer reflect its native-like conformation. This could affect downstream ion activation exper-iments, leading to erroneous conclusions about the native-like structure. One such activation strategy is surface-induced dis-sociation (SID), which generally causes native-like protein complexes to dissociate along the weakest subunit interfaces, re-vealing critical information about the complex’s native-like topology and subunit connectivity. If the quaternary structure has been disturbed then the SID fingerprint will shift as well. Thus, SID was used diagnose source-induced quaternary struc-ture rearrangement and help tune an instrument’s source and other upstream transmission regions to strike the balance be-tween signal intensity, adduct removal, and conserving the native-like structure. Complementary to SID, electron-capture dissociation (ECD) can also diagnose rearranged quaternary structures and was used after in-source activation to confirm that the subunit interfaces were rearranged, opening the structure to electron capture and subsequent dissociation. These results provide a valuable guide for new practitioners to native mass spectrometry and highlight the importance of using standard protein complexes when tuning new instrument platforms for optimal native mass spectrometry performance.
Chemistry
What problem does this paper attempt to address?
The problem that this paper attempts to solve is in native mass spectrometry (nMS), how to maintain the native conformation of biomolecular complexes while increasing signal intensity and removing adducts, and avoid quaternary structure rearrangement caused by improper instrument tuning. Specifically, when the ion source of the mass spectrometer is tuned to maximize signal intensity or remove adducts, it may cause the tertiary and quaternary structures of the biomolecular complex to be rearranged, so that it no longer reflects its native conformation. This rearrangement will affect subsequent ion activation experiments, leading to wrong conclusions about the native conformation. To diagnose and avoid this structure rearrangement, the authors used surface - induced dissociation (SID) and electron capture dissociation (ECD) as diagnostic tools. SID can reveal the dissociation path of protein complexes along the weakest subunit interface, providing key information about the native topology and subunit connection of the complex. If the quaternary structure is perturbed, the SID fingerprint will also change accordingly. Therefore, SID can be used to diagnose source - caused quaternary structure rearrangement and help tune the source of the instrument and other upstream transmission regions to balance the requirements of signal intensity, adduct removal and maintaining the native conformation. ECD induces fragmentation through interaction with low - energy electrons, producing c/z - type fragments, which usually come from flexible and solvent - exposed regions, thereby providing information about the structure of native protein complexes in the gas phase. ECD can also be used to diagnose source - caused structure changes. Overall, by using SID and ECD, this paper aims to provide guidance for new native mass spectrometry users, emphasizing the importance of using standard protein complexes when tuning new instrument platforms to achieve optimal native mass spectrometry performance.