Optimizing Native Ion Mobility Q-TOF in Helium and Nitrogen for Very Fragile Noncovalent Interactions

Valérie Gabelica,Sandrine Livet,Frédéric Rosu
DOI: https://doi.org/10.1007/s13361-018-2029-4
2018-07-16
Abstract:The meaningful comparison of ion mobility (IM) results and of collision cross section (CCS) values on different platforms is a prerequisite for using CCS for identification or structural assignment. The amount of internal energy imparted to the ions prior to the ion mobility cell is a source of experimental variation. Here we investigated the effects of virtually all tuning parameters of the Agilent 6560 IM-Q-TOF on the arrival time distributions of Ubiquitin7+, and found conditions in which the native state prevails. We will discuss the effects of solvent evaporation conditions in the source, in the entire pre-IM DC voltage gradient, and with the funnel RF amplitudes, and will also report on ubiquitin7+ conformations in different solvents, including native supercharging conditions. Collision-induced unfolding (CIU) can be conveniently provoked in two distinct regions: behind the source capillary (by changing the fragmentor voltage) and in the trapping funnel (by changing the trap entrance grid delta voltage). The softness of the instrumental conditions were then optimized with the benchmark DNA G-quadruplex [(dG4T4G4)2.(NH4+)3-8H]5-, for which ion activation results in ammonia loss. To reduce the ion internal energy and obtain the intact 3-NH4+ complex, we reduced the post-IM voltage gradient, but this resulted in a lower IM resolving power due to increased diffusion behind the drift tube. The article thus describes the various trade-offs between ion activation, ion transmission, and ion mobility performance for native MS of very fragile structures.
Quantitative Methods,Biomolecules
What problem does this paper attempt to address?
The problem that this paper attempts to solve is how to optimize the operating conditions of ion mobility (IM) - quadrupole time - of - flight mass spectrometry (Q - TOF) in a helium and nitrogen environment in a mass spectrometer, in order to reduce the destruction of very fragile non - covalent structures, especially when analyzing the native states of these complex molecules. Specifically, the researchers focused on the following points: 1. **Reduce ion activation in the pre - IM region**: By adjusting various parameters of the instrument (such as voltage gradient, radio - frequency amplitude, etc.), reduce the activation before ions enter the IM cell, thereby maintaining the native conformation of ions. 2. **Optimize the conditions in the post - IM region**: By reducing the voltage gradient in the post - IM region, reduce the further activation of ions after the IM cell, especially for very sensitive nucleic acid complexes, such as G - quadruplexes. 3. **Balance ion transmission efficiency and softness**: During the optimization process, it is necessary to maintain high ion transmission efficiency while minimizing the internal energy of ions as much as possible to avoid unnecessary desolvation or depolymerization. 4. **Compare the effects under different solvent conditions**: The effects of different solvent conditions (such as water/methanol and 0.2% formic acid vs. 1% acetic acid) on ion conformation were studied to determine the conditions that can best maintain the native state. Through these optimizations, the researchers hope to be able to analyze and characterize these very fragile non - covalent structures more accurately, especially when conducting collision - induced unfolding (CIU) experiments, and to better understand the behavior of these molecules in the gas phase.