Gas-liquid Two-phase Flow Measurement of Throat-Extended Venturi Meter Based on Differential Pressure Ratios and Fluctuations

Peng Suo,Jiangtao Sun,Shiying Shi,Shijie Sun,Feng He,Xiaokai Zhang,Mengxian Shen,Xiaolin Li,Jinfei Li,Lijun Xu
DOI: https://doi.org/10.1109/tim.2024.3481576
IF: 5.6
2024-01-01
IEEE Transactions on Instrumentation and Measurement
Abstract:This article presents a novel approach to measuring gas-liquid two-phase flow using a throat-extended Venturi meter (TEVM) based on differential pressure (DP) ratios and fluctuations. The over-reading (OR) and mass quality are the dominant parameters for accurate mass flowrate measurement by a TEVM. First, a liquid OR model is established based on multiple DP ratios, gas-to-liquid density ratio, and liquid densitometric Froude number. This model is solved by integrating a deep neural network (DNN) with an iterative optimization technique. Subsequently, a mass quality model is developed based on multi-DP fluctuations and other related parameters, which can also be solved using a DNN. Finally, the liquid and gas mass flowrates are obtained sequentially from the preceding results. The proposed method addresses the issue of significant errors in the direct solution of the conventional dual-parameter measurements. It utilizes both the static and dynamic features of multi-DP to achieve the simultaneous measurement of gas and liquid mass flowrate using a single TEVM. The experiment was conducted within the a gas flowrate range of 1.3×10 -4 ~5.1×10 -3 kg/s, and a liquid flowrate range of 1.12~3.95 kg/s. The results demonstrate that the relative error for liquid and gas mass flowrate is within ±3% and ±10%, respectively. This indicates that the proposed method is a cost-effective and viable solution for gas-liquid two-phase flow metering using a TEVM, and it can be further adapted to other experimental conditions.
What problem does this paper attempt to address?