Finite Element Method-Based Hybrid MRI/CBCT Generation to Improve Liver Stereotactic Body Radiation Therapy Targets Localization Accuracy

Zeyu Zhang,Mark Chen,Ke Lu,Dongyang Guo,Zhuoran Jiang,Hualiang Zhong,Jason K. Molitoris,Phuoc T. Tran,Fang-Fang Yin,Lei Ren
DOI: https://doi.org/10.1109/trpms.2024.3466184
2024-01-01
IEEE Transactions on Radiation and Plasma Medical Sciences
Abstract:Cone-beam CT (CBCT) is commonly used in treatment imaging, but its limited soft tissue contrast presents challenges for liver tumor localization. As a result, indirect localization methods relying on the liver’s boundary are commonly utilized, which have limited accuracy for tumor localization. On-board MRI offers superior soft tissue contrast but is limited by the cost. To address this, we devised a method to generate onboard virtual MRI by integrating pretreatment MRI with onboard CBCT, enhancing liver SBRT tumor localization accuracy. We employed a finite element method (FEM) for deformable mapping, deforming prior liver MR images onto CBCT geometry to create a virtual MRI. This hybrid virtual-MRI/CBCT (hMRI-CBCT) approach was evaluated in a pilot study involving 48 patients. The hMRI-CBCT demonstrated superb soft-tissue contrast with clear tumor visualization. Registration accuracy of hMRI-CBCT to planning CT significantly surpasses the onboard CBCT to planning CT registration, particularly for tumors not near the liver boundary, with an average error reduction of 1.53 ± 2.16 mm. Our study demonstrated that hybrid MRI/CBCT can apparently reduce localization errors in liver SBRT, potentially improving tumor control and reducing toxicities, and opening avenues for further margin reduction and dose escalation.