Measurement of Atmospheric Neutron Spectrum at Near Space Altitudes Using a CLYC(Ce) Detector

Weixin Zhou,Xiaoxin Zhang,Jianguang Guo,Guanglin Yang,Weiguo Zong,Qinglong Yu,Yigang Yang
DOI: https://doi.org/10.1109/tim.2024.3449976
IF: 5.6
2024-01-01
IEEE Transactions on Instrumentation and Measurement
Abstract:To provide the reference neutron radiation data for the single-event effect (SEE) analysis and dose equivalent calculation, a compact and wide energy range CLYC(Ce) (Cs2LiYCl6 : Ce3+) neutron detector was designed. CLYC(Ce) scintillator has the ability to detect fast neutrons, thermal neutrons, and gamma rays simultaneously and can be used in various radiation scenes. The detector was successfully mounted on a high-altitude balloon for near-space neutron measurement, and the measured results showed that charged particles, neutrons, and gamma rays reached the maximum flux at an altitude of about 20 km, called the Regener-Pfotzer (R-P) maximum phenomenon. The neutron deposition energy spectrum measured by the detector during the balloon flight was also used to unfold the atmospheric neutron energy spectrum in the range of 0.1-100 MeV at an average altitude of 27 km. These measured results can provide references for radiation protection and single-particle effects in the near-space environment.
What problem does this paper attempt to address?