Deep Learning-Aided Respiratory Motion Compensation in PET/CT: Addressing Motion Induced Resolution Loss, Attenuation Correction Artifacts and PET-CT Misalignment.

Yihuan Lu,Fei Kang,Duo Zhang,Yue Li,Hao Liu,Chen Sun,Hao Zeng,Lei Shi,Yumo Zhao,Jing Wang
DOI: https://doi.org/10.1007/s00259-024-06872-x
2024-01-01
Abstract:Respiratory motion (RM) significantly impacts image quality in thoracoabdominal PET/CT imaging. This study introduces a unified data-driven respiratory motion correction (uRMC) method, utilizing deep learning neural networks, to solve all the major issues caused by RM, i.e., PET resolution loss, attenuation correction artifacts, and PET-CT misalignment. In a retrospective study, 737 patients underwent [18F]FDG PET/CT scans using the uMI Panorama PET/CT scanner. Ninety-nine patients, who also had respiration monitoring device (VSM), formed the validation set. The remaining data of the 638 patients were used to train neural networks used in the uRMC. The uRMC primarily consists of three key components: (1) data-driven respiratory signal extraction, (2) attenuation map generation, and (3) PET-CT alignment. SUV metrics were calculated within 906 lesions for three approaches, i.e., data-driven uRMC (proposed), VSM-based uRMC, and OSEM without motion correction (NMC). RM magnitude of major organs were estimated. uRMC enhanced diagnostic capabilities by revealing previously undetected lesions, sharpening lesion contours, increasing SUV values, and improving PET-CT alignment. Compared to NMC, uRMC showed increases of 10
What problem does this paper attempt to address?