Electrochemical Detection of Pb(II) Ions in Aqueous Solution by Carbon Nanospheres Modified Boron-Doped Diamond Electrode

Nan Wang,Jiaqi Li,Yitian Ma,Ahua Xu,Huajin Tao,Baoyu Huang,Song Hao,Jianjun Liao,Shiwei Lin,Yipeng Tang,Ying Hou,Chao Bian,Xiaogan Li
DOI: https://doi.org/10.1007/s41664-024-00320-3
2024-01-01
Journal of Analysis and Testing
Abstract:In this paper, a sensor using carbon nanospheres modified boron-doped diamond (CNs/BDD) electrode for detecting heavy metal ions rapidly in aqueous solution is studied. The CNs/BDD electrode is characterized systematically by techniques including scanning electron microscopy (SEM), X-ray diffractometer (XRD) and X-ray photoelectron spectroscopy (XPS). The electrochemical properties of the CNs/BDD electrode, BDD electrode and Au electrode are compared by cyclic voltammetry. It is found that the CNs/BDD electrode has more negative hydrogen evolution potential, a wider potential window and a larger effective electrode area, which is conducive to the accurate detection of heavy metal ions. The square-wave anodic stripping voltammetry (SWASV) is applied to assess and quantify the sensing capability of the CNs/BDD electrode. The effects of deposition potential, deposition time and other experimental parameters on the detection performance are studied. Under the optimal experimental conditions, CNs/BDD electrodes exhibit remarkable detection capabilities for lead (Pb(II) ions, in which a linear relationship within the concentration range from 1 to 140 µg/L with a 0.9994 correlation coefficient is obtained. Meanwhile, it has excellent repeatability with a low relative standard deviation (RSD) of 0.57
What problem does this paper attempt to address?