Development of an Endotracheal Tube Position Detector

Wei Pan
2004-01-01
Abstract:Misled esophageal intubation can bring about a serious accident during operations and emergence treatments if it is not recognized immediately. In order to detect the position of endotracheal tube, a magnetic field strength detector has been designed using Hall-effect sensor. Before intubation, a stylet with a magnet on its tip is inserted into the endotracheal tube and the Hall sensor is fixed on the skin of neck over the glottis. When the magnet, fixed on the tip of the stylet in the tube, passes through the Hall sensor, the position of magnet would be detected, as a result, the position of endotracheal tube would be known. Clinical tests have indicated that this detector is stable and has high accuracy. So it may have promising applications both in endotracheal intubation training and clinical utilization.
What problem does this paper attempt to address?