[Magnetic induced phase shift detection system based on a novel sensor for cerebral hemorrhage]

Jie Liu,Lian Yan,Mingxin Qin,Haisheng Zhang,Mingsheng Chen
DOI: https://doi.org/10.7507/1001-5515.202305048
2024-06-25
Abstract:The main magnetic field, generated by the excitation coil of the magnetic induction phase shift technology detection system, is mostly dispersed field with small field strength, and the offset effect needs to be further improved, which makes the detection signal weak and the detection system difficult to achieve quantitative detection, thus the technology is rarely used in vivo experiments and clinical trials. In order to improve problems mentioned above, a new Helmholtz birdcage sensor was designed. Stimulation experiment was carried out to analyze the main magnetic field in aspects of intensity and magnetic distribution, then different bleeding volume and bleeding rates experiments were conducted to compared with traditional sensors. The results showed that magnetic field intensity in detection region was 2.5 times than that of traditional sensors, cancellation effect of the main magnetic field was achieved, the mean value of phase difference of 10 mL rabbit blood was (-3.34 ± 0.21)°, and exponential fitting adjusted R 2 between phase difference and bleeding volumes and bleeding rates were both 0.99. The proposed Helmholtz birdcage sensor has a uniform magnetic field with a higher field strength, enable more accurate quantification of hemorrhage and monitored change of bleeding rates, providing significance in magnetic induced technology research for cerebral hemorrhage detection.
What problem does this paper attempt to address?