Abstract:Real-time continuous monitoring of brain edema can significantly reduce mortality and disability after craniocerebral injury. By combining the magnetic induction phase shift (MIPS) method and computer programming, a real-time continuous monitoring system for brain edema was established with advantages including non-invasive and non-contact monitoring capabilities. After the injection of 6 ml of simulated brain edema solution, the mean MIPS amplitude change was (-7.9743 +/- 0.9338)degrees. Compared with previous work, the sensitivity was greatly improved. Thirty rabbits (24 in the experimental group and 6 in the control group) were used for animal experiments. In the experimental group, the MIPS values showed an obviously decreasing trend with an average change of (-3.9438 +/- 1.6392)degrees, (-7.3206 +/- 1.6598)degrees and (-10.9686 +/- 1.9897)degrees after 6, 12 and 24 hours, respectively; the brain water content (BWC) increased with time. In contrast, MIPS values of rabbits in the control group showed no obvious upward or downward trend within 24 hours, with an average change of (-0.1730 +/- 2.08602)degrees. The BWC and MIPS measurements between the control group and the experimental group had significant differences at different times. These results indicate that the method developed in this study is capable of continuously monitoring gradual increases in the brain edema volume.