Contamination of Microbiological Aerosol Generated by Pathogenic Microbiological Labs

Dong Xiao
2013-01-01
Abstract:Objective To simulate the microbiological aerosol generated by the pathogenic microbiological lab,and to provide scientific data on risk evaluation and protection measures for lab assistants.Methods Serratia marcescens was used as the indicator microorganism.Many experimental operations and accidental cases were simulated in the lab,such as mix-up,accidental fall-off,centrifugal operation,centrifugal tube break,injection infection,anatomization of animals and leakage of a biolsafety cabinet.The risk of microbiological aerosol was analyzed by quantitative air sampling and natural agar sedimentation methods.Results The concentration of microbiological aerosol was different under laboratory conditions.The highest intensity of the microbiological aerosol of mix-up was 1409 CFU/m3,accidental fall-off was 9346 CFU/m3,centrifugal operation was 10 CFU/m3,centrifugal tube break was 138 CFU/m3,injection infection was more than 8232 CFU/m3,anatomization of animals was 4 and biological safety cabinet leakage was above 801 CFU/m3,respectively.Conclusion The intensity of the microbiological aerosol generated by pathogenic microbiological labs varies widely.A high concentration of microbiological aerosol can be generated in a local environment.Experimental skill training and application of primary respiratory protection equipments should be emphasized.