Abstract 16746: Lower Fluoroscopy Frame Rate During Device Implantation Reduces Radiation Exposure

John Duell,Hiroko Beck,Stephen J. Synowski,Benjamin Remo,Thomas Klein,Anastasios Saliaris,Vincent See,Ramarao Vunnam,Rui Huang,Steven R. Shorofsky,Timm Dickfeld
IF: 37.8
2013-01-01
Circulation
Abstract:Introduction: Radiation exposure is related in a dose-dependent fashion to long-term risk of a variety of malignancies for both patients and operators in the electrophysiology laboratory. All attempts to reduce the dose without compromising care should be taken. Hypothesis: We examined whether decreasing fluoroscopy frame rate from 3 frames per second (FPS) to 2 FPS during permanent pacemaker (PPM) and implantable cardiac defibrillator (ICD) implantation can safely reduce radiation exposure without a detrimental impact in clinical outcomes. Methods: Single or dual chamber device implantations performed at a single tertiary center from July 2011 to June 2013 were identified and enrolled into the study. Average fluoroscopy dose, fluoroscopy duration, case length, major complication rate, and procedure-related mortality were compared retrospectively. Complications were defined as any event requiring re-accessing the pocket, removal of hardware, additional procedures or medication changes. Results: Out of a t...