The Establishment of Using Liquid Chip Technology to Quantify Prostate Specific Antigen in Human Serum

Xiaomei Yan
2009-01-01
Abstract:Objective To quantify prostate specific antigen in human serum using liquid chip method and evaluate the performance of the method.Methods Liquid chip method was employed to detect the PSA in 50 cases of benign prostatic hyperplasia patients'serum,evaluating the index of linear range,detection limit and accuracy,and make a correlation analysis with chemiluminescence immunoassay(CLIA).Results The linear range for the measurement of PSA was 0.022~129.6ng/mL and the lowest detection limit for PSA was 0.018 ng/ml.The intra-and inter-assay coefficients of variation were 2.18%~2.28% and 1.61%~4.18% respectively.Liquid chip method and CLIA were used to quantify PSA in human serum.The results showed that there were no significant differences in these two methods(P0.05),but they had a good correlation(r=0.9984).Conclusion Liquid chip immunoassay has wide linear range,high sensitivity,repeatability,saving time and samples,with potential clinical applications.
What problem does this paper attempt to address?