Highly Sensitive Detection Ofl-Lactate and Pyruvate by Amperometric Flow-Injection Analysis Based on Enzymatic Substrate Recycling and Selective Detection of Hydrogen Peroxide

Tao Yao,Masahiro Satomura,Tuketoshi Nakahara
DOI: https://doi.org/10.1002/elan.1140070418
IF: 3.077
1995-01-01
Electroanalysis
Abstract:The poly(1,2‐diaminobenzene)film‐coated platinum electrode was used as an amperometric detector in a flow‐injection system with the lactate dehydrogenase/lactate oxidase coimmobilized reactor involving amplification by substrate recycling. Both L ‐lactate and pyruvate are recycled enzymatically during their passage through the enzyme reactor in the presence of reduced nicotinamide adenine dinuceotide (NADH) and oxygen in the carrier stream. As a result, a large amount of hydrogen peroxide is generated in the enzyme reactor. The poly(1,2‐diaminobenzene)‐coated electrode could only selectively detect the hydrogen peroxide generated in the enzyme reactor, without an increase in the base‐line current owing to the direct oxidation of NADH, because the poly(1,2‐diaminobenzene) film effectively prevented NADH in the carrier stream from reaching the electrode surface and allowed only the hydrogen peroxide to penetrate into the polymerized film. In the present flow‐injection analysis (FIA) system, both L ‐lactate and pyruvate were determined with a 400‐fold increase in sensitivity compared with the unamplified responses. The detection limit was 2 × 10 −9 M (20 fmol) for 10 μL sample injection.
What problem does this paper attempt to address?