Comparison of Fetal Heart Rate Baseline Estimation by the Cardiotocograph Network and Clinicians: A Multidatabase Retrospective Assessment Study

Xiuyu Pan,Yue Lu,Mei Zhong,Huijin Wang,Zheng Zhang,Jieyun Bai
DOI: https://doi.org/10.21203/rs.3.rs-1793588/v1
2022-01-01
Abstract:Objective This study aims to compare the fetal heart rate (FHR) baseline predicted by the cardiotocograph network (CTGNet) with that estimated by clinicians. Results A total of 1267 FHR recordings acquired with different electrical fetal monitors (EFM) were collected from five datasets. Each FHR baseline was estimated by clinicians and CTGNet, respectively. And agreement between CTGNet and clinicians was evaluated using the kappa statistics, intra-class correlation coefficient, and the limits of agreement. The number of differences < 3 beats per minute (bpm), 3–5 bpm, 5–10 bpm and ≥ 10 bpm, is 64.88%, 15.94%, 14.44% and 4.74%, respectively. Kappa statistics and intra-class correlation coefficient are 0.873 and 0.969, respectively. Limits of agreement are − 6.81 and 7.48 (mean difference: 0.36 and standard deviation: 3.64). An excellent agreement was found between CTGNet and clinicians in the baseline estimation from FHR recordings with different signal loss rates.