Influencing Factors Affecting the Results of Multi-Tumor Marker Protein Chip Detection During Image Analysis

WU Xiaodong
2012-01-01
Abstract:Objective To improve the accuracy rate of multi-tumor marker(MTM) protein chip detection method and provide more reliable evidence for early clinical diagnosis of tumors.Methods Three groups of images were selected randomly,including 10 normal sample images,10 positive sample images and 10 strong positive sample images.The images of each group were detected by direct-measurement method(DMM) and segmentation-measurement method(SMM),respectively,and the results were statistically compared and analyzed.Results No significant difference was found in the normal group and positive group measured by the two methods(P0.05).For the strong positive group,the difference was also not significant in the strong positive markers selves and those markers nonadjacent to strong positive markers(P0.05).However,significant difference was found in the markers adjacent to strong positive markers measured by the two methos(P0.05).Conclusions From our study,we could conclude that segmentation-measurement method was more accurate than direct-measurement method in tumor marker detection using multi-tumor marker protein chip.Thus,image segmentation processing should be made during detecting MTM with protein chip detection system.
What problem does this paper attempt to address?