Comparative Study Between Bio-chip Detection and Chemiluminescent Immunoassay for Multi-tumor Markers

陆汉魁,季红,朱瑞森,辛丽君,顾建英
DOI: https://doi.org/10.3969/j.issn.1006-1703.2005.02.010
2005-01-01
Abstract:Bio-chip system for multi-tumor marker detection system (Chip-C-12) can simultaneously measure 12 tumor markers: AFP, CEA, PSA, fPSA, CA125, CA15-3, CA19-9, CA242, NSE, β-HCG, HGH and Ferritin. This study aims to investigate the feasibility and clinical utilities of Chip-C-12, as compared with automated chemiluminescent immunoassay (ACLIA). 22 serum samples from normal subjects and 138 serum samples from various cancer patients were both assayed by Chip-C-12 and ACLIA. The results showed that the quantitative values for 12 tumor markers determined by Chip-C-12 were significantly related with those from ACLIA (r=0.32 for normal,r=0.28 for cancers,P0.01). For cancer samples, the average single positive rate was 77.5% by Chip-C-12 vs. 82.1% by ACLIA (P0.05). The results suggested that Chip-C-12 may be defined as an easy and applicable method for screening tumor markers for clinical diagnosis and management; help increase single diagnostic sensitivity for cancer patients. However, its quantitative accuracy needs further evaluation.
What problem does this paper attempt to address?