The Potential Ability of Plan Complexity Metrics on the Dose Calculation and Plans Delivery in Intensity Modulated Radiotherapy

S. Li,H. Luo,X. Tan,Y. Wang,F. Jin
DOI: https://doi.org/10.1016/j.ijrobp.2023.06.1546
2023-01-01
Abstract:This study is the first overall assessment of the impact of various complexity metrics on the accuracy of TPS calculation and Linac delivery. Of the metrics studied, MIs and MIa metrics have a standout impact on the ability of the TPS calculation and delivery system, extra attention should be paid during the planning process. It is inappropriate to utilize calculation-based QA to predict the results of measurement-based QA since there is a poor correlation between the two. Furthermore, calculation-based QA outperforms measurement-based QA in identifying highly complex plans, which can further guide clinical QA process optimization and save limited clinical resources.