ION CHROMATOGRAM METHOD FOR ANALYZING ANIONS AND CATIONS IN PORE WATER FROM MARINE SEDIMENTS

Lan Ge,Yang Tiezhao,Jin Yang,Shao‐Yong Jiang
2006-01-01
Abstract:We use ion chromatogram technique to determine the anion and cation ion concentrations in pore waters.Metrosep A Supp4-250 type column and 1.8 mmol/L Na_(2)CO_(3)+1.7 mmol/L NaHCO_(3)mixed solution in anion system are used to determine concentrations of Cl~(-),Br~(-)and SO~(2-)_4 in pore water,and Metrosep C 2-150 type column and 4 mmol/L tartaric acid+0.75 mmol/L C_(7)H_(5)NO_(4) mixed solution in cation system are used to determine concentrations of Na~(+),NH~+_4,K~(+),Ca~(2+)and Mg~(2+).The results indicate that correlation coefficient between measured and standard values for each ion in standard solutions preponderates over 0.999.The detection limit is less than 0.5 mg/L,and the relative standard deviation is better than 3%.In particular,we carried out special experiment to measure Br~(-)concentration under high Cl~(-) concentration background as well as NH~+_4concentration under high Na~(+) concentration background.Right now,a tentative ion chromatograph method for analyzing concentrations of cations and anions in pore water from marine sediments has been established in our lab.This method has been proved to be rapid,simple,sensitive,and accurate.
What problem does this paper attempt to address?