The Relations among Inspired or Expired Isoflurane Concentration and Anaesthesia Duration Or/and Vaporizer Output Concentration During Low-Flow Closed Circuit Anaesthesia

Yanling Zhu
1999-01-01
Abstract:Objective To evaluate the relations among inspired or expired isoflurane concentration (Ci, Ce) and anaesthesia duration (T ) or/and vaporizer output concentration (Co ) during lowflow closed circuit anaesthesia(LFCCA). Methods Under intravenous combined inhalational anaesthesia, 43 ASA gradeⅠ or Ⅱ patients undergoing elective surgical procedure,were randomly divided into 3 groups according to the different Co,the Co of group Ⅰ,Ⅱ or Ⅲ was 1. 5%,3. 0% or 5. 0% respectively. The fresh oxygen flow was fixed at 0.5L· min~(-l). Simultaneously,propofol was continuously administered via vein to achieve adequate anaesthesia depth. Datex monitor was applied to measuring Ci and Ce. The coeffecients and regression equations among Ci or Ce and T and /or Co,were worked out by least square method. Results Ci and Ce did not equilibrate during 240-min inhalational anesthesia. The ratio of Ce/Ci increased following anaesthesia. There was linear correlation between Ci or Ce and T ;The higher Co,the higher sloope of the equation,meaning Ci and Ce increased more rapidly. There also was linear correlation among Ci or Ce and T and Co. Conclusion The prolongation of LFCCA duration requires lowering properly the vaporizer output concentration in order to prevent deep anesthesia,