Various Effects on Axial Resolution in Fiber-Optical Confocal Microscopy System

Ding Xue-mei
2008-01-01
Infrared and Laser Engineering
Abstract:For the Axial Resolution improvement purpose,fiber spot size,pupil radius,pupil filters,matching of fiber and lens were studied in this paper.Based on the three-dimension transfer function of Fiber-optical Confocal Microscopy System (FOCSM),the axial response function was set up and the Various Effects on Axial Resolution were analyzed.For the axial-resolution improvement,the experiment result indicates that parameter A related to fiber profile spot should be less than 3,and make the numerical aperture of the lens smaller than that of fiber-optic.However the axial-resolution cannot benefit from the pupil filters.
What problem does this paper attempt to address?