Abstract:Objective To evaluate the feasibility of endovascular repair of the aortic arch using a triple branched stent graft in a novel in vitro vascular model.Methods A triple branched stent graft was designed for aortic arch and its three branch vessels.A novel in vitro vascular model was established to test this stent graft.Attempts were made to optimize the procedure of stent grafting and to evaluate the feasibility of this device.Results The new type of in vitro vascular model was manufactured successfully.The triple stent graft above was compressed and crimped well to fit into the corresponding delivery system and was deployed successfully in the in vitro vascular model,with its main body and branches positioned properly.The secure fixation of the stent graft was achieved in this model.Conclusions We design and produce a new type of branched stent graft for aortic arch.This system can deliver the stent grafts to the proper position of the target vessel.The in vitro vascular model can simulate the real endovascular circumstance in human body and meets the requirement for testing the stent graft above-mentioned.