Impact of a Novel Whole-Heart Motion-Correction Algorithm on Non-Contrast Cardiac CT Images and Coronary Artery Calcium Scores

Mengzhen Wang,Jing Fan,Min Yang,Le Qin,Fuhua Yan,Wenjie Yang
DOI: https://doi.org/10.1007/s42058-023-00125-w
2023-01-01
Chinese Journal of Academic Radiology
Abstract:To evaluate the impact of a new whole-heart motion-correction algorithm (SnapShot Freeze 2, SSF2) on non-contrast cardiac CT images and the coronary artery calcium scores (CACS). 100 consecutive patients were prospectively included to undergo ECG-gated CT CACS scanning. Images were reconstructed with both standard (STD) algorithm and SSF2. CACS was calculated by a semi-automatic software. Image noise and signal-to-noise ratio (SNR) of the aortic root were measured. Two radiologists evaluated the subjective image quality of coronary arteries and calcified plaques using a 5-point scale and 3-point scale, respectively. CACS of 278 vessels with calcified plaques in 100 patients were calculated. The overall median (range) Agatston score, volume score, and mass score of SSF2 were 172.3 (307.9), 138 (252.9) and 38.3 (65.6), respectively, lower than 193.4 (342.0), 160.0 (284.8) and 39.3 (74.4) of STD (all P < 0.001). SSF2 provided lower noise (17.4 ± 2.7 VS. 18.5 ± 2.9 HU, P < 0.001) and higher SNR (2.90 ± 0.58 VS. 2.70 ± 0.55, P < 0.001) than those of STD. The subjective image quality scores of SSF2 group were significantly higher than that of STD group (all P < 0.001). The new whole-heart motion-correction algorithm improves non-contrast cardiac CT image quality, lowers CACS, and shows a potential for increasing CACS accuracy.
What problem does this paper attempt to address?