Simultaneous determination of polysorbate 20 and polyethylene glycol 4000 in paliperidone palmitate injection by HPLC-ELSD

MA Yong-chao,JI Wei,XUE Ying,XIA Yun-qing,ZHOU Feng-mei,LIU Wan-hui,XU Li-xiao
DOI: https://doi.org/10.16155/j.0254-1793.2023.03.15
2023-01-01
Abstract:Objective: To establish an HPLC-ELSD method for the simultaneous determination of Polysorbate(Tween) 20 and Polyethylene Glycol(PEG) 4000 in Paliperidone palmitate injection. Methods: The separation was performed on a TSK gel 3000 PWxl(7.8 mm×300 mm) column with 20 mmol·L -1 ammonium acetate aqueous solution and acetonitrile(60∶40) as the mobile phase by isocratic elution. And the flow rate was 0.6 mL·min -1 . The column temperature was 25 ℃ and the sampling volume was 20 μL. The ELSD was maintained at 50 ℃ throughout the determination. The nebulizer gas flow rate was 2.0 L·min -1 (0 ℃,1 atm). Results: PEG4000 showed good linearity within the range of 0.3-0.75 mg·mL -1 (r=0.999 4). Tween 20 showed good linearity within the range of 0.12-0.3 mg·mL -1 (r=0.999 6). The minimum detection limit of PEG4000 and Tween 20 were 0.02 mg·mL -1 ,0.05 mg·mL -1 , respectively. The minimum quantitative limit of PEG4000 and Tween 20 were 0.03 mg·mL -1 , 0.07 mg·mL -1 , respectively. The repeatability RSD of PEG4000 and Tween 20 were 2.3% and 1.9%, respectively. The average recoveries of PEG4000 was 98.7% and the RSD was 0.90%(n=6). The average recovery of Tween 20 was 96.3% and the RSD was 0.80%(n=6). Conclusion: The method for simultaneous determination of Tween 20 and PEG4000 in Paliperidone palmitate injection by using HPLC-ELSD is precise and suitable.