Research of Nanosecond Pulse Current Detection Technology

WU Bin,CHEN Yonggang,BAO Jiali,CHEN Xinhua
DOI: https://doi.org/10.3969/j.issn.1674-1633.2023.05.006
2023-01-01
Abstract:Objective The clinical application of nanosecond pulsed electric field treatment equipment requires accurate and stable acquisition of pulse output current amplitude. Due to the steep front edge, short pulse width, and strong electromagnetic interference of therapeutic pulses, a new pulse current detection technology has been developed to achieve the pulse current detection function.Methods A high-performance isolated current sensor was used for pulse current sampling. After the current signal was modulated and attenuated, it was transmitted through impedance matching and differential methods. The field programmable gate array(FPGA)cached the current digital signal after analog-to-digital conversion, and finally it was transmitted to a microcontroller chip through communication for analysis and processing of current data. A multi-layer printed circuit board(PCB) design was adopted to increase anti-interference. Results The static direct current test and dynamic pulse test error under actual load were less than 5%, and the stable sampling at the nanosecond level is achieved. Conclusion The nanosecond pulse current detection module can meet the application requirements of medical grade treatment equipment and is easy to integrate, with strong clinical practicality.