Determination of pentachlorophenol in breast milk by immunoaffinity column coupled with high performance liquid chromatography-tandem mass spectrometry

HAN Mu-Ke,YIN Jie,WEN Kai,SHAO Bing
DOI: https://doi.org/10.19812/j.cnki.jfsq11-5956/ts.2023.07.026
2023-01-01
Abstract:Objective To construct an analytical method for the determination of pentachlorophenol residues in human breast milk by immunoaffinity column purification coupled with high performance liquid chromatography-tandem mass spectrometry(HPLC-MS/MS). Methods The human breast milk samples were extracted with 5%triethylamineacetonitrile-water(7:3, V/V), specifically cleaned by immunoaffinity column, with methanol containing 0.05% ammonium fluoride and ultrapure water containing 0.05% ammonium fluoride as mobile phases, gradient elution, separated by ACQUITY UPLC HSS T3 column, detected in multiple reaction monitoring(MRM) mode, and quantifed by internal standards. Results Pentachlorophenol showed a good linear relationship in the range of 0.02-20.00 μg/L, and the correlation coefficient reached 0.9998. The recoveries was 98.72%-106.95%, the relative standard deviations were less than 6.09%, and the matrix effects was 94.45%-101.61% under the 3 different spiked concentrations of low, medium and high. The limit of the detection was 0.008 μg/L and the limit of quantification was 0.02 μg/L. The immunoaffinity column could be reused at least 8 times under experimental conditions, reducing the experimental cost by more than 87%. Conclusion This method specifically cleans up samples based on the immunoaffinity column, which has good reproducibility and high accuracy, which can effectively isolate and detect pentachlorophenol residues in human breast milk, and can be applied to the qualitative and quantitative detection of pentachlorophenol in human breast milk samples.
What problem does this paper attempt to address?