Diffusion-weighted imaging of the scrotum:readout-segmented echo-planar imaging versus single-shot echo-planar imaging
Jie Cai,Liang Wang,Ming Deng,Xiangde Min,Liang Li,Chaoyan Feng,Basen Li,Zan Ke,Peipei Zhang
DOI: https://doi.org/10.3760/cma.j.issn.1005-1201.2016.07.007
2016-01-01
Abstract:Objective The purpose of this study was to compare the image quality of readout-segmented EPI (RS-EPI) and that of standard single-shot echo-planar imaging (SS-EPI) in patients with scrotal diseases. Methods Initial diagnosis of scrotal diseases were included in the prospective study, all patients underwent scrotum routine MRI and RS-EPI, SS-EPI sequence at 3.0 T. A total of 38 patients were recruited qualitative assessment ,and 29 patients proved by operation and pathology (malignant 21 cases and benign 8 cases) were included quantitative analysis.For qualitative comparison of image quality, two readers independently assessed the two sets of DWI, which consisted of identification of structure and geometric distortion of scrotum (epididymis,testes, lesion) using a 5-point Likert scale. For assessment of image parameters, signal-to-noise ratio (SNR), contrast, contrast-to-noise ratio (CNR),apparent diffusion coefficient value of normal testis and testis distortion ratio of anatomical(T2WI)/DWI fusion image were calculated. Inter-observer agreement was assessed with the Kappa statistics, the differences of scores of <br> RS-EPI and SS-EPI were compared by using Wilcoxon rank-sum test, and image parameters were compared by using paired sample t test. Results There was good reader agreement in the scores, the Kappa value were 0.77, 0.74, 0.80, 0.87. The difference of identification of structure and geometric distortion on RS-EPI and SS-EPI had statistic significance, the RS-EPI was superior to SS-EPI in image quality (all P<0.05).The SNR, contrast of benign lesion for two sequences had no statistic significance(P values were 0.352, 0.124, respectively), but the difference of CNR had statistic significance(P<0.05). The SNR,contrast and CNR of malignant lesion for two sequences had statistic significance (all P<0.05). The mean size of the testis was (35.61 ± 9.78)mm, the mean distance in fusion image of RS-EPI or SS-EPI with T2WI were (3.80 ± 1.32)mm and (7.54 ± 2.62)mm, and the mean distortion ratios of the two set were (11.1 ± 0.6)% and (22.4 ± 19.2)%, respectively, the difference of two sets had statistic significance(P<0.05). There was no significant difference in the ADC obtained by using either DWI method(t=0.396,P=0.796), the mean ADC values of normal testes for RS-EPI and SS-EPI, respectively, were (1.18 ± 0.06) × 10-3mm2/s, (1.24 ± 0.11) × 10-3mm2/s.Conclusion RS-EPI DWI images is a feasible technique in the scrotum for producing high-resolution DWI with reduced geometric distortion and offers potentially superior image quality compared to SS-EPI at 3.0 T.