Automatic Multi-Plug Filtration Cleanup Tip-Filtration with Ultra-Performance Liquid Chromatography/Tandem Mass Spectrometry Detection for 22 Pesticide Residues in Typical Vegetables

Lili Ma,Lihua Cao,Yuechao Feng,Li Jia,Cong Liu,Qi Ding,Jia Liu,Peng Shao,Canping Pan
DOI: https://doi.org/10.1093/chromsci/bmac104
2023-01-01
Journal of Chromatographic Science
Abstract:An automatic multi-plug filtration cleanup (m-PFC) tip-filtration method was developed to reduce the manual operation workload in sample preparation. In this work, m-PFC was based on multi-walled carbon nanotubes mixed with primary secondary amines and anhydrous magnesium sulfate (MgSO4) in a packed column for analysis of pesticide residues followed by ultra-performance liquid chromatography coupled with tandem mass spectrometry. Method validation was performed on 22 pesticide residues in carrot, spinach and leek, at spiked levels of 5, 10 and 50 μg/kg, respectively. The average recoveries were between 70.1 and 119.5% with associated relative standard deviations <20% (n = 6) indicating satisfactory accuracy and repeatability. Matrix-matched calibration curves were performed with the correlation coefficients (R2) higher than 0.9903 within a linearity range of 5-100 ng/mL. The limits of quantification were 5 μg/kg for all the pesticides in carrot, spinach and leek matrices. The developed method was successfully used to determine pesticide residues in market samples.
What problem does this paper attempt to address?