Sol-gel dip-coated TiO<sub>2</sub> nanofilms reduce heat production in titanium alloy implants produced by microwave diathermy

Zheng Wei,Ziwei Zhang,Xianxuan Feng,Yun Cai,Jiajia Yang,Zikai Hua,Yuehong Bai,Yiming Xu
DOI: https://doi.org/10.1080/02656736.2022.2152500
2023-01-01
International Journal of Hyperthermia
Abstract:Objective: To verify that the TiO2 nanofilm dip-coated by sol-gel can reduce titanium alloy implants (TAI)'s heat production after microwave diathermy (MD). Methods: The effect of 40W and 60W MD on the titanium alloy substrate coated with TiO2 nanofilm (Experimental Group) and the titanium alloy substrate without film (Control Group) were analyzed in vitro and in vivo. Changes in the skeletal muscle around the implant were evaluated in ex vivo by histology. Results: After 20 min of MD, in vitro the temperature rise of the titanium substrate was less in the Experimental Group than in the Control Group (40W: 1.4 degrees C vs. 2.6 degrees C, p < .01, 60W: 2.5 degrees C vs. 3.7 degrees C, p<.01) and in vivo, the temperature rise of the muscle tissue adjacent to TAI was lower in the Experimental Group than in the Control Group (40W: 3.29 degrees C vs. 4.8 degrees C, p < .01, 60W: 4.16 degrees C vs. 6.52 degrees C, p < .01). Skeletal muscle thermal injury can be found in the Control Group but not in the Experimental Group. Conclusion: Sol-gel dip-coated TiO2 nanofilm can reduce the heat production of TAIs under single 40(similar to)60W and continuous 40W MD and protect the muscle tissue adjacent to the implants against thermal injury caused by irradiation. [GRAPHICS] .