Research on the Correction Method of Reflected Energy in Radiation Temperature Measurement

Yanxiu Wei,Bojun Sun,Xiaogang Sun,Kaixin Xue
DOI: https://doi.org/10.1016/j.infrared.2022.104378
IF: 2.997
2022-01-01
Infrared Physics & Technology
Abstract:Measure the target temperature with a radiation thermometer, the noise of the ambient radiance reflected by the target can cause measurement errors, especially when the ambient temperature is higher than the target temperature. In this paper, the one-point reflection correction method (OPRCM) is proposed based on the rela-tionship between the band emissivity and the blackbody radiance corresponding to the target temperature. According to this method, the target's emissivity needs not be known, the instrument only needs to be calibrated in full-range at one ambient temperature, and the target temperature can be accurately measured under different ambient temperatures, with the measurement range being-10 degrees C similar to 50 degrees C, the measurement accuracy is better than +/- 1 degrees C. Therefore, it is of great significance to improve the accuracy of radiation temperature measurement.
What problem does this paper attempt to address?