Comparison of Compressed Sensing-Sensitivity Encoding (CS-SENSE) Accelerated 3D T2W TSE Sequence Versus Conventional 3D and 2D T2W TSE Sequences in Rectal Cancer: a Prospective Study

Xiaoling Gong,Daguang Wen,Hong Wei,Yu Shen,Yujiao Deng,Ya Wang,Mingtian Wei,Xiaoxiao Zhang,Xiaoyong Zhang,Ziqiang Wang,Bing Wu
DOI: https://doi.org/10.1007/s00261-022-03636-9
IF: 2.4
2022-01-01
Abdominal Radiology
Abstract:This study aimed to evaluate the image quality and diagnostic value of compressed sensing-sensitivity encoding (CS-SENSE) accelerated 3-dimensional (3D) T2-weighted turbo spin-echo (T2W TSE) sequence in patients with rectal cancer compared with conventional 3D and 2-dimensional (2D) sequences. A total of 54 patients who underwent the above three sequences were enrolled. Two radiologists independently reviewed the image quality using an ordinal 5-point Likert scale. The quantitative measurement was performed to calculate the signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). The diagnostic value was assessed using TN staging, extramural vascular invasion and mesorectal fascia status. Friedman and McNemar’s tests were applied for comparative analysis. Forty-two patients were successfully included. Compared with 3D and 2D sequences, the CS-SENSE 3D sequence speeded up by 39
What problem does this paper attempt to address?