Quantifying 224ra/228th Disequilibrium in Sediments Via a Pulsed Ionization Chamber (PIC)

Wen Liu,Chunqian Li,Pinghe Cai,Shibin Zhao,Jinjia Guo,William C. Burnett,Shasha Song,Guangquan Chen,Bochao Xu
DOI: https://doi.org/10.1016/j.marchem.2022.104160
IF: 3.994
2022-01-01
Marine Chemistry
Abstract:A new method for measuring( 224)Ra in sediments is established by use of a pulsed ionization chamber (PIC). Based on the difference in their half-lives,Rn- 220 can be distinguished from Rn-222 so that one can easily measure Ra-224 and its parent( 228)Th in sediment samples. Results show that the measurement efficiency is 0.22 +/- 0.02 cpm/dpm, and is not readily affected by moisture. At an optimum airflow rate of 1.2 L/min, the precision of the( 224)Ra and Th-228 measurements is estimated at +/- 8%. Based on controlled experimental field application and a field application, we show that the Ra-224/Th-228 disequilibrium in sediments can be assessed and both 224Ra deficits and ingrowth to equilibrium may be quantified. The PIC device is simple to use, inexpensive and portable, making it especially suitable for determinations of sediment samples in the field.
What problem does this paper attempt to address?