Abstract:The cantilever beam photoacoustic(PA) spectroscopy technology has attracted the attention of many scholars due to its high sensitivity. For different application objects, the background gas type of the gas to be measured may be different. At present, there are few reports on the influence of background gas types on the intensity of PA signal for cantilever enhanced PA gas detection technology. In this paper, we carried out CO detection under the background gases of nitrogen (N 2 ), sulfur hexafluoride (SF 6 ), helium (He) and argon (Ar), respectively. The results show that for CO gas with the same concentration, the order of PA signal intensity from strong to weak is Ar>He >N 2 >SF 6 . Through the theoretical derivation of the intensity of PA signal, the theoretical formula of the intensity of PA signal is obtained. When the detection device and the gas to be tested remain unchanged, the intensity of PA signal is related to the heat capacity ratio of the gas, and the greater the heat capacity ratio, the stronger the PA signal.
What problem does this paper attempt to address?