An Integrated Microfluidic System for Multi-Target Biochemical Analysis of a Single Drop of Blood.

Zhi-Qiang Zuo,Jian-Zhang Pan,Qun Fang
DOI: https://doi.org/10.1016/j.talanta.2022.123585
IF: 6.1
2022-01-01
Talanta
Abstract:Herein, we described an integrated microfluidic system for multi-target biochemical analysis of micro-volumes of blood samples. This system mainly consists of a microfluid handling module, an integrated microwell array chip, a temperature control module and a spectral detection module. A novel microfluidic plasma extraction approach was developed by coupling the filter membrane-based plasma separation technique with the microfluidic liquid handling technique. With this approach, quantitative extraction of trace blood with volumes as low as a drop (30 mu L) can be automatically realized. Multiple operations in multi-target biochemical analysis are achieved including micro-volume blood collecting, quantitative plasma extraction, plasma dilution, plasma distribution, transferring, biochemical reaction and absorption spectroscopy detection. This system was applied to the multi target biochemical analysis of glucose, cholesterol and total protein in blood samples and could achieve the determination of the target analytes of a 30-mu L blood sample within 11 min. The experimental results are consistent with those obtained by a commercial biochemical analyzer.
What problem does this paper attempt to address?