Validation of a Novel Renal Denervation System With Cryoablation

Meng Ji,Han Chen,Li Shen,You Zhang,Zhifeng Yao,Yizhe Wu,Xu Yawei,Junbo Ge
DOI: https://doi.org/10.1016/j.jacbts.2021.11.002
2022-01-01
Abstract:• With the concept of cryoablation, we innovatively applied liquid nitrogen in RDN and designed a dedicated balloon catheter and system for Cryo-RDN. • In a swine model, this Cryo-RDN system demonstrated device-related safety and efficacy. The extent and depth of nerve ablation were efficient and stable. Sustained decreases in renal and serum norepinephrine also suggested effective ablation in the long term. • Cryo-RDN also met safety endpoints in 6 patients with resistant hypertension. The 24-hour ambulatory blood pressure and office blood pressure of all 6 patients were reduced after 6 months compared with their baseline values, providing clinical support for further large-scale clinical studies. Recently, we designed a renal denervation with cryoablation (Cryo-RDN) system using liquid nitrogen and proved its short-term safety and effectiveness. In this study, we first conducted a 6-month follow-up in a swine model. Renal sympathetic nerve activity remained at a significantly lower level than that of the control group after 6 months. In patients with resistant hypertension, Cryo-RDN demonstrated preliminary safety. Renal function fluctuations and vascular-related complications were not detected. In addition, the average 24-hour systolic and diastolic blood pressure decreased by 12.17 ± 8.35 mm Hg and 8.50 ± 3.83 mm Hg at the 6-month follow-up, respectively, compared with their baseline values.