Study of <SUP>99m</SUP>Tc absorption on micro-sized ion exchange resins to achieve high activity for SPECT

Qingyang Wei,Zhenlei Lyu,Nianming Jiang,Zhihan Deng,Tianpeng Xu,Lifeng Sun,Xuefeng Zhang,Zuo-Xiang He,Yaqiang Liu,Tianyu Ma
DOI: https://doi.org/10.1016/j.apradiso.2022.110256
IF: 1.787
2022-01-01
Applied Radiation and Isotopes
Abstract:In single-photon emission computed tomography (SPECT), a micro-sized Tc-99m source is routinely used for performance measurement, geometry calibration, and system matrix generation. Therefore, a micro-sized source is critical in nuclear instrument production and quality control. Standard methods can only produce a point source with a large size and low total activity, as they are limited by the concentration of the Tc-99m solution. The absorption of Tc-99m on ion exchange resins has been used; however, few studies have quantitatively evaluated the absorption process and optimized the source activity. This paper proposes a procedure for producing a microsized Tc-99m resin source with a super-high concentration, as well as a method for the fast measurement of the point source time-activity curve (TAC). Experiments on two resin point sources with diameters of 0.681 mm and 0.326 mm were carried out. Two semi-empirical models, including the first kinetic model and the pseudo-second-order rate equation model, were used to fit TACs. The results show the first kinetic model fit better, which suggests an acquisition time of 2-4 h is needed for optimization. The verification experiment demonstrates a resin point source with a diameter of 0.35 mm and total activity of 10.6 mCi (i.e., 59.1 Ci/mL concentration) was produced.
What problem does this paper attempt to address?