Preliminary Study on Radioluminescence Dose Rate Measurement of Α-Al 2O 3:C Coupled to SiPM

Kaiyong Tang,Haijun Fan,Hui Cui,Haijun Li,Siyuan Zhang,Li Fu
DOI: https://doi.org/10.1016/j.radmeas.2022.106851
IF: 1.743
2022-01-01
Radiation Measurements
Abstract:In this work the first investigation into the use of a alpha-Al2O3:C detector directly coupled to a SiPM as a real-time dosimeter is presented. Three kinds of dosimeters with a single-chip detector with the size of 4 x 4 x 0.8 mm(3), 320 pieces of alpha-Al2O3:C detectors with the size of 4 x 4 x 0.8 mm(3) and a single crystal alpha-Al2O3:C detector with the size of 13 x 13 x 20 mm(3) were tested. The average net counts rates during irradiation at a constant dose rate of 100 mu Gy/h are 40.3, 415.1 and 858.9 counts/s/(mu Gy/h) for the three kinds of dosimeters respectively. The RL sensitivity of alpha-Al2O3:C with the size of 4 x 4 x 0.8 mm3 is about 27 times that of LiF:Mg,Cu,P with the same size. The three dosimeters show a linear response within the dose rate range from 10 mu Gy/h to 100 mGy/h, from 1 mu Gy/h to 5 mGy/h and from 0.35 mu Gy/h to 500 mu Gy/h respectively. The lower limit of the dose rate measurement by the dosimeter with the single crystal alpha-Al2O3:C detector is the lowest. The dosimeter based on a single crystal alpha-Al2O3:C detector and a SiPM is particularly attractive and can be applied for the dose rate measurement as low as 0.35 mu Gy/h. The dosimeter may have the potential to replace G-M counter tube with a low range for personal and environmental dose monitoring.