Renewable-surface graphite-ceramic glucose biosensor

杨绍明,林贤福
DOI: https://doi.org/10.3969/j.issn.1000-0720.2003.z1.185
2003-01-01
Abstract:A surface-renewable glucose biosensor is reported. The glucose biosensor is developed using glucose oxidase (GOD) encapsulated in organically modified solgel glass (ORMOSIL) network in the composite matetial. The organic group in the ORMOSIL network controls the hydrophobicity of the electrode surface and thus limits the wettability of the electrode surface. The graphite powder provides the conductivity for the electrode.Ferrocenecarboxylic acid in phosphate buffer solution (pH 7.0) transfers electron between enzyme and electrode. Cyclic volammetry and amperometric measurements have been used to exmine the electrochemical behavior of glucose biosensor as shown in Fig. 1 and Fig.2. The electrode gives a linear response range of 1 -20mM glucose with a sensitivity of 3.26 μA· mM-1. The electrode can be renewed easily in reproducible manner by a simple polishing step.
What problem does this paper attempt to address?