Evaluation of SO42-,NO3- and NH4+ measurements using ambient PM2.5 real-time monitoring instruments

C. Yuan,Tao Wang,X.M. Gao,W. Nie,P.K.K. Louie,Z. Xu,Wenxing Wang
IF: 4.236
2012-01-01
Environmental Chemistry
Abstract:Field studies were conducted in Hong Kong to assess two different real-time techniques for measuring SO2-4,NO-3 and NH+4 in PM2.5 by comparison with a honeycomb denuder filter-pack system.An ambient ion monitor(AIM,URG9000B) showed good performance for NO-3 and NH+4,but obvious overestimation for SO2-4,which was attributed to the positive interference by high ambient SO2 concentrations which could not be totally eliminated in the parallel denuder.To overcome this problem,a series of experiments were conducted,which lead to an improved system including two denuders working in series and with the use of a new denuder liquid(5 mmol · L-1 H2O2 +5 mmol · L-1 NaOH).The new system could absorb high concentration of SO2(even up to 260 μg · m-3) so that the SO2-4 measurement results would not be affected.Another monitor for measuring aerosols and gases(MARGA,Model ADI 2080),developed by ECN,Metrohm and Applikon,was also evaluated.MARGA showed excellent performance for the measurements of SO2-4 and NH+4,but exhibited high NO-3 measurement results.The accuracy of NO-3 measurements should be evaluated further.This study demonstrated the need for evaluating new measurement techniques in different atmospheric environments to determine the accuracy and precision of the measurements.
What problem does this paper attempt to address?