MO-F-16A-05: Spot-Scanning Proton Therapy Patient-Specific Quality Assurance: Our Methodology and Results From 295 Treatment Plans

Dennis Stephen Mackin,Xiang Zhang,Yupeng Li,H Li,Richard Wu,Falk Poenisch,Kazumichi Suzuki,M Kerr,Charles Holmes,X.R. Zhu,Narayan Sahoo,Michael Gillin
DOI: https://doi.org/10.1118/1.4889176
IF: 4.506
2014-01-01
Medical Physics
Abstract:Purpose: To report on the patient-specific quality assurance (PSQA) results for 295 spot-scanning proton therapy treatment plans from the MD Anderson PTC-Houston. We show how the results differed by treatment site and how they were affected by the treatment plan optimization method and by a range shifter in the treatment field. We also discuss some causes of PSQA problems. Methods: The PSQA procedure, which is designed to verify both the accuracy of the treatment planning system's (Eclipse™ v8.9) dose calculations and the dose delivery of the Hitachi PROBEAT synchrotron, consists of (1) an end-to-end test in which the beam is delivered and measured at the prescribed gantry angle, and (2) additional dose plane measurements made from gantry angle 270°. HPlusQA™ software automatically performs the gamma analysis with criteria 3% (dose tolerance), 3 mm (distance-to-agreement, DTA) and 2%, 2 mm. Passing is defined as at least 90% of the pixels having a gamma score less than 1. Results: The PSQA gamma passing rate was 96.2% for 3%, 3 mm, and 85.3% for 2%, 2 mm. The rate depended on the treatment site. For example, the 3%, 3 mm passing rate was 95% for head and neck plans, vs 100% for prostatemore » plans. The passing rates of multi- vs. single-field optimization plans did not significantly differ. However, the rate for fields with range shifters was 94.8±0.6%, vs 99.0±0.6% for those without (p = 0.002). Longitudinal dose gradients caused most of the low scores. Overestimation of the calculated dose proximal to the spread-out Bragg peak (SOBP) caused many of the others. Conclusion: The planned and delivered doses consistently agreed within tolerance levels. Minor dose modeling deficiencies remain proximal to the SOBP. The 3% dose tolerance, 3 mm DTA, with 90% pixel passing rate is a reasonable action level for 2D gamma comparisons.« less
What problem does this paper attempt to address?