APPLICATION OF LOW-TEMPERATURE VACUUM CONCENTRATION IN AUTOMATED ON-LINE MONITORING OF GROSS ALPHA AND BETA IN WATER

Shuijun He,J.T He,Manchun Liang,Guofeng Su,Ke Li
DOI: https://doi.org/10.1299/jsmeicone.2019.27.1258
2019-01-01
Abstract:An automated on-line monitoring method for gross alpha and beta in water based on low-temperature vacuum concentration and liquid scintillation counting analysis (LSC) measurement is proposed, in order to solve the low-level automation problems of artificial measurement of gross alpha and beta radiation in water, which results in high dependence on human being and impossibility to be used in large area. In order to improve the stability of recovery rate and recovery rate in vacuum cryogenic concentration, magnetic stirring was used to prevent the loss caused by explosion boiling and splash. The experimental results show that the recovery rate of this method is stable, up to 90%, and the speed of concentration can reach 10ml/min, which can greatly shorten the pretreatment time of LSA measurement. The minimal detectable activity concentration (MDC) of total alpha and beta can be achieved in far less than 0.5Bq/L and 1Bq/L respectively by using the existing low background small LSA within one hour of measurement time. This method successfully solves the problem of unmanned on-line monitoring of total alpha and beta in water, and can also be applied to on-line monitoring of other nuclides in water, which will greatly improve the level of on-line monitoring of nuclides.