Analysis of the problems existed in the quality control of HER-2 gene testing in Guangdong Province
Shuhua LI,Xiaoke ZHENG,Fang WANG,li LI,Shijun SUN,Dan HE,Yan PENG,Jianwen ZHOU,Qiong HE,Yuxin ZHU,Liantang WANG,Zunfu KE
DOI: https://doi.org/10.3969/j.issn.1674-6929.2018.02.008
2018-01-01
Abstract:Objective Relying on the molecular pathology group of Guangdong province clinical quality control center to carry out the external quality control for HER-2 gene detection by fluorescence in situ hybridization (FISH) methods in Guangdong province hospitals, in order to understand the standard operating procedure(SOP)key node of HER-2 gene detection entirely and better guide the targeted therapy of breast can-cer. M ethods The paraffin sections with different positive patterns of HER-2 were selected as the quality control tablets,and the cases of the same expression pattern were performed consecutively,and the first and last 2 pieces were tested to ensure the consistency of quality control tablets.Each participating laboratories adopted the FISH methods of their own work and selected the corresponding probes and digestive enzymes according to the actual situation. The quality control results were interpreted by quality control team members in a double-blind manner. R esults From the results of 26 laboratories participating in the quality control of HER-2 gene detection,all reagents used in the laboratories were approved by China Food and Drug Administration,and had no significant effect of these 3 probes on the quality control results. There were still some details of operation and interpretation in some individual laboratories. As a unit mistaken for cluster amplification, the self-agree-ment results of specimen①was 96.15%,specimen②was 84.62%,and the consistency of specimen③was up to 100%.The specimens were digested with Pepsin and proteinase K,and there were no obvious difference be-tween the 2 kinds of enzyme digestion.Meanwhile,microscopic examinations showed that DAPI staining were unclear in 2 units, it suggested the existence of excessive digestion. C onclusion Each laboratory should strictly follow the operating procedures of each laboratory to ensure that the temperature, buffer and cleaning liquid concentration and pH value,and probe configuration are accurate in the dyeing process.Only by constant-ly improving the external quality control system, the scientific design basis and clinical experiment process steps and evaluation standards,cooperating with the inspection team who knows the content,and regularly mon-itoring activities,the accuracy and consistency of HER-2 detection can be ensured in our province.