Multi-channel respiratory signal detection system for 4D-CT in radiotherapy by measuring the back pressure

Yuan Zheng,Yahui Peng,Haizhen Yue,Haiyan Xiang,Yi Du
DOI: https://doi.org/10.1109/EMBC46164.2021.9631091
2021-01-01
Abstract:This study proposes a novel respiratory signal detection system for 4D-CT in radiotherapy by measuring back pressure changes at multiple positions on CT couch. The 12-channel pressure sensor is fixed on CT couch to obtain patient's back pressure signal. The 12-channel signal is transmitted to a PC at a sampling rate of 50 Hz after a signal conditioning circuit and an analog-digital converter. The amplitude of pressure changes is characterized to select the optimal channel. This system is validated by comparing with the respiratory signal collected synchronously with a real-time position management (RPM) system on 10 healthy volunteers. The correlation coefficient between the signals is 0.82 +/- 0.09 (standard deviation) and the time shift is 0.32 +/- 0.15 second. We conclude that the back pressure signal acquired by the proposed system has the potential to replace the clinical RPM system for respiratory signal detection in 4D-CT data acquisition.