Preparation And Characterization Of Nm-Sized Electrochemical Sensors

Yh Shao,Mv Mirkin,G Fish,S Kokotov,D Palanker,A Lewis
1996-01-01
Abstract:The nanometer-size glass-sealed metal ultramicroelectrodes (UMEs) have been prepared using a laser-based micropipette puller. The tip was exposed to solution either by etching of a small portion of glass insulator or by micropolishing. The characterization of the UMEs was carried out by combination of steady-state voltammetry, scanning electron microscopy (SEM), and scanning electrochemical microscopy (SECM). The cyclic voltammograms (CV) obtained have a regular shape with very small capacitive and resistive background. The effective electrode radii obtained from voltammetry were between 2 and 500 nm. From the SEM micrographs, the shape of polished tips appears to be close to a microdisk, while the geometry of unpolished electrodes is closer to conical. Accordingly, the SECM current-distance curves (i(T) - d) obtained with polished electrodes fit well the theory for a disk-shaped tip, while a 20-nm-radius unpolished electrode was shown to be a fairly sharp cone with a height-to-radius ratio of about 2.5. A submicrometer-size ion selective electrode was prepared by coating an unpolished Ag tip with silver iodide and protective Nafion film.
What problem does this paper attempt to address?