Three-Dimensional In Vivo Imaging Of Human Anterior Eye Segment With Optical Coherence Tomography

Jiaxu Hong,Xiaoyu Liu,Wen Wen,Yubo Fan,Xinghuai Sun,Jianjiang Xu
2011-01-01
Abstract:Objects: A novel computer method for three-dimensional in vivo imaging of human anterior eye segment is described with optical coherence tomography. Methods: A laser beam was projected at an oblique angle on the cornea and scanned to acquire 18 optical section images at 10-degree intervals across the corneal area. Tissue's edges were detected using Canny detector, and Track Points arithmetic was designed to extract the single-pixel edges, which were smoothed by curve fitness. According to the characters of rotating image, which appeared displaced because of the motion blur of patients, the classical hat-head method was improved to build 3D model for OCT images sequence. Three B-Spine interpolation and curve fitness would be used to obtain the outlines if they were lost in original OCT images. The optical section images were segmented to construct a series of anterior eye. Imaging was performed in normal subject and patients. Results: A series of digital 3D model of anterior eye was reconstructed, which allowed enhanced visualization of the micro-structures, including the section of anterior eye segment at any rotating degree, the contours of corneal thickness and curvature, and the depths and angles of anterior chamber. Conclusion: A method for in vivo three-dimensional anterior eye imaging was developed, which has potential as a tool for objective, noninvasive and real-time evaluation of diseases.