Force distribution within the frame of self-expanding transcatheter aortic valve: Insights from in-vivo finite element analysis

Tian-Yuan Xiong,Elisa Stoppani,Matthieu De Beule,Fei Chen,Yi-Jian Li,Yan-Biao Liao,Yuan Feng,Peter de Jaegere,Mao Chen
DOI: https://doi.org/10.1016/j.jbiomech.2021.110804
IF: 2.4
2021-01-01
Journal of Biomechanics
Abstract:We sought to assess the amount and distribution of force on the valve frame after transcatheter aortic valve replacement (TAVR) via patient-specific computer simulation. Patients successfully treated with the self-expanding Venus A-Valve and multislice computed tomography (MSCT) pre- and post-TAVR were retrospectively included. Patient-specific finite element models of the aortic root and prosthesis were constructed. The force (in Newton) on the valve frame was derived at every 3 mm from the inflow and at every 22.5° on each level. Twenty patients of whom 10 had bicuspid aortic valve (BAV) were analyzed. The total force on the frame was 74.9 N in median (interquartile range 24.0). The maximal force was observed at level 5 that corresponds with the nadir of the bioprosthetic leaflets and was 9.9 (7.1) N in all patients, 10.3 (6.6) N in BAV and 9.7 (9.2) N for patients with tricuspid aortic valve (TAV). The level of maximal force located higher from the native annulus in BAV and TAV patients (8.8 [4.8] vs. 1.8 [7.4] mm). The area of the valve frame at the level of maximal force decreased from 437.4 (239.7) mm2 at the annulus to 377.6 (114.3) mm2 in BAV, but increased from 397.5 (114.3) mm2 at the annulus to 406.7 (108.9) mm2 in TAV. The maximum force on the bioprosthetic valve frame is located at the plane of the nadir of the bioprosthetic leaflets. It remains to be elucidated whether this may be associated with bioprosthetic frame and leaflet integrity and/or function.
What problem does this paper attempt to address?